| Specification |
Information |
| Working (Operating) Temperature |
15-70°C |
| Storage Temperature |
0-70°C |
| Temperature Resolution |
0.000001°C |
| Reproducibility/Repeatability |
0.1% RSD |
| Resolution |
0.0001 MJ/Kg |
| Results per hour |
10 samples per hour using 2 oxygen bomb vessels |
| Measuring range max |
99MJ, 99,000,00 Joules |
| Working temperature min |
1°C |
| Working temperature max |
50°C |
| Temperature Measurement Resolution |
10ppm (parts per million) |
| Cooling Medium |
Air, External Cooler |
| Type of Cooling |
Airflow, Ambient Air |
| Oxygen Operating Pressure Max |
40 bar |
| Balance/Scale Interface |
RS232, 1200 to 38400 Baud (settable) |
| Printer Interface |
RS232, 1.2Kb to 115.2Kb |
| Computer Interface |
USB or RS232 |
| Calorimeter Power Input |
36W/12VDC/ 3Amps |
| Interface External Keyboard |
PS2 |
| Oxygen Filling |
Manual |
| De gasification |
Manual |
| Halogen (Decomposition) Vessel |
Yes, optional |
| Analysis according to DIN 51900 |
Yes |
| Analysis according to ASTM D240 |
Yes |
| Analysis according to ASTM D4809 |
Yes |
| Analysis according to ASTM D5865 |
Yes |
| Analysis according to ASTM E711 |
Yes |
| Analysis according to ISO 1928 |
Yes |
| Dimensions |
27cm (W) x 27cm (L) x 24cm (H) |
| Weight |
9kg |
| Permissible Ambient Temperature |
1-35°C |
| Permissible Relative Humidity |
80% |
| RS232 Interface |
Yes, 2 x RS232 |
| USB Interface |
Yes |
| Mains Voltage |
220-240/ 100-120V |
| Mains Frequency |
50/60 Hz |