| Specification |
Information |
| Laboratory Working Operating Conditions (Temperature) |
15-30°C |
| Storage Temperature |
0-70°C |
| Reproducibility/Repeatability |
0.1% RSD |
| Resolution |
0.0001 MJ/Kg |
| Results per hour |
Up to 4-6 samples per hour |
| Measuring range max |
99MJ, 99000J, 99KJ/g |
| Working temperature min |
5°C |
| Working temperature max |
50°C |
| Temperature Measurement Resolution |
10ppm (parts per million) |
| Cooling Medium |
Air, External Cooler |
| Filling Oxygen Operating Pressure Max |
40 bar |
| Balance / Scale Interface |
RS232 |
| Printer Interface |
RS232 |
| PC Interface |
USB or RS232 |
| Power Input |
12W |
| Interface External Keyboard |
PS2 |
| Oxygen Filling |
Manual |
| De-gasification |
Manual |
| Halogen (Decomposition) Vessel |
Yes, optional |
| Analysis according to DIN 51900 |
Yes |
| Analysis according to ASTM D240 |
Yes |
| Analysis according to ASTM D4809 |
Yes |
| Analysis according to ASTM D5865 |
Yes |
| Analysis according to ASTM E711 |
Yes |
| Analysis according to ISO 1928 |
Yes |
| Dimensions |
350mm x 280mm x 240mm |
| Weight |
~ 8kg |
| Permissible Ambient Temperature |
10-35°C |
| Permissible Relative Humidity |
80% |
| RS232 Interface |
Yes |
| USB Interface |
Yes |
| Voltage |
220-240/ 100-120V, 12VDC, 1Amp |
| Frequency |
50/60 Hz |