ASTM D5865-12
STANDARD TEST METHOD FOR GROSS CALORIFIC VALUE OF COAL AND COKE
Scope
This test method pertains to the determination of the gross calorific value of coal and coke by either an isoperibol or adiabatic bomb calorimeter. The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard. This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use. Specific hazard statements are given in Section 8.
TERMINOLOGY
For additional definitions of terms used in this test method, refer to Terminology D121.
Adiabatic Calorimeter - a calorimeter that operates in the adiabatic mode and may or may not use a microprocessor. The initial temperature before initiating the combustion and the final temperatures are recorded by the operator or the microprocessor.
Automated Calorimeter - a calorimeter which has a microprocessor that takes the thermometric readings and calculates the Calibration Value and the Heat of Combustion Values.
British Thermal Unit (BTU) - is the amount of heat required to raise the temperature of one pound - mass (lbm) of liquid water at one atmospheric pressure one degree Fahrenheit at a stated temperature. The results of combustion calorimetric tests of fuel for steam power plants may be expressed in terms of the 1956 International Steam Table calorie (I.T. cal) which is defined by the relation, 1 I.T. cal = 4.1868 J. The BTU used in modern steam tables is defined by the means of the relation, 1 I.T. cal/g = 1.8 I.T. BTU/lb. Thus, 1 I.T. BTU/lb = 2.326 J/g.
Calorific Value - the heat produced by combustion of a unit quantity of a substance under specified conditions.
Calorimeter - a device for measuring calorific value consisting of a bomb, its contents, a vessel for holding the bomb, temperature measuring devices, ignition leads, water, stirrer, and a jacket maintained at specified temperature conditions.
Gross Calorific Value (gross heat of combustion at constant volume), Qv (gross) - the heat produced by complete combustion of a substance at constant volume with all water formed condensed to a liquid.
Heat of Formation - the change in heat content resulting from the formation of 1 mole of a substance from its elements at constant pressure.
Isoperibol Calorimeter - a calorimeter that operates in the isoperibol mode and uses a microprocessor to record the initial and final temperatures and make the appropriate heat leak corrections during the temperature rise. It determines when the calorimeter is in equilibrium and ignites the sample and determines when the calorimeter has reached equilibrium after ignition.
Net Calorific Value (net heat of combustion at constant pressure), Qp (net) - the heat produced by combustion of a substance at a constant pressure of 0.1MPa (1 atm), with any water formed remaining as vapor.
Corrected Temperature Rise - the calorimeter temperature change caused by the process that occurs inside the bomb corrected for various effects.
Heat Capacity - the energy required to raise the temperature of the calorimeter one arbitrary unit.
Discussion - The heat capacity can also be referred to as the energy equivalent or water equivalent of the calorimeter.
SUMMARY OF TEST METHOD
The heat capacity of the calorimeter is determined by burning a specified mass of benzoic acid in oxygen. A comparable amount of the analysis sample is burned under the same conditions in the calorimeter. The calorific value of the analysis sample is computed by multiplying the corrected temperature rise, adjusted for extraneous heat effects, but the heat capacity and dividing by the mass of the sample.
Oxidation of coal after sampling can result in a reduction of calorific value. In particular, lignite and sub-bituminous rank coal samples may experience greater oxidation effects than samples of higher rank coals. Unnecessary exposure of the samples to the air for the time of sampling or delay in analysis shall be avoided.
SIGNIFICANCE AND USE
The gross calorific value can be used to compute the total calorific content of the quantity of coal or coke represented by the sample for payment purposes.
The gross calorific value can be used for computing the calorific value versus sulfur content to determine whether the coal meets regulatory requirements for industrial fuels.
The gross calorific value can be used to evaluate the effectiveness of beneficiation processes.
The gross calorific value can be required to classify coals according to Classification D388.
APPARATUS AND FACILITIES
Test Area - An area free from drafts, shielded from direct sunlight and other radiation sources. Thermostatic control of room temperature and controlled relative humidity are desirable.
Combustion Bomb - Constructed of materials that are not affected by the combustion process or the products formed to introduce measurable heat input or alteration of end products. The bomb shall be designed so that all liquid combustion products can be completely recovered by washing the inner surfaces. There shall be no gas leakage. The bomb shall be capable of withstanding a hydrostatic pressure test to 20 MPa (3000 psig) at room temperature without stressing any part beyond its specified elastic limit.
Balance - a laboratory balance capable of weighing the analysis sample to the nearest 0.0001g. The balance shall be checked weekly, at a minimum, for accuracy.
Calorimeter Vessel - Made of metal with a tarnish-resistant coating, with all outer surfaces highly polished. Its size shall be such that the bomb is completely immersed in water during determination. A stirrer shall be provided for uniform mixing of the water. The immersed portion of the stirrer shall be accessible to the outside through a coupler of low thermal conductivity. The stirrer speed shall remain constant to minimize any temperature variations due to stirring. Continuous stirring for 10 min shall not raise the calorimeter temperature more than 0.01C when starting with identical temperatures in the calorimeter, test area and jacket. For calorimeters having a bucket it can be a separate component of the bomb. The vessel shall be of such construction that the environment of the calorimeter's entire outer boundaries can be maintained at a uniform temperature.
Jacket - a container with the inner perimeter maintained at constant temperature ± 0.1°C (isoperibol) or at the same temperature ± 0.1°C as the calorimeter vessel (adiabatic) during the test. To minimize convection, the sides, top and bottom of the calorimeter vessel shall not be more than 10mm from the inner surface of the jacket. Mechanical supports for the calorimeter vessel shall be of low thermal conductivity.
Thermometers - Platinum resistance or thermistor thermometers - shall be capable of measuring to the nearest 0.0001°C. These types of thermometers consist of two major subsystems. The first and most obvious is the temperature sensing probe itself. The second and equally important aspect is the measurement subsystem. For both subsystems, the relationship between the thermometer resistance and temperature shall be well characterized. The absolute temperature shall be known to ± 0.1°C at the temperature of the calorimetric measurement.
Assessing the valid working range of the calorimeter, as outlined in 10.9, is sufficient to demonstrate that all important aspects of the calorimeter functionality, including the thermometry, are in good working order. The traceability of the heat of combustion of the benzoic acid calibrant in addition to meeting the aforementioned criteria.
Liquid-in-Glass Thermometers - Conforming to the requirements for thermometers S56C, S116C, or S117C as prescribed in Specification E2251.
Thermometer Accessories - A magnifier is required for reading liquid-in-glass thermometers to one tenth of the smallest scale division. The magnifier shall have a lens and holder designed so as to minimize errors as a result of parallax.
Sample Holder - An open crucible or platinum, quartz, or base metal alloy. Before use in the calorimeter, heat treat base metal crucibles for a minimum of 4 hours at 500°C to ensure the crucible surface is completely oxidized. Base metal alloy crucibles are acceptable, if after three preliminary firings, the weight does not change by more than 0.0001g.
Ignition Fuse - Ignition fuse of 100mm length and 0.16mm (No.34 B&S gauge) diameter or smaller. Nickel-chromium alloy (Chrome C) alloy, cotton thread, or iron wire are acceptable. Platinum or palladium wire, 0.10mm diameter (No.38 B&S gauge), can be used provided constant ignition energy is supplied. Use the same type and length (or mass) of ignition fuse for calorific value determinations as used for standardization.
Ignition Circuit - A 6-30V alternating or direct current is required for ignition purposes. A step-down transformer connected to an alternating current circuit, capacitors, or batteries can be used. For manually operated calorimeters, the ignition circuit switch shall be of the momentary double-contact type, normally open except when held closed by the operator. An ammeter or pilot light can be used in the circuit to indicate when current is flowing.
Controller - For automated calorimeter, capable of charging the bomb; filling the calorimeter vessel; firing the ignition circuit; recording calorimeter temperature before, during and after the test; recording the balance weights; and carrying out all necessary corrections and calculations.
Crucible Liner - Quartz fiber or alundum for lining the crucible to promote complete combustion of samples that do not burn completely during the determination of the calorific value.
REAGENTS
Reagent Water - Conforming to conductivity requirements for Type II of Specification D1193 for preparation of reagents and washing of the bomb interior.
Purity of Reagents - Use reagent grade chemicals conforming to the specification of the Committee on Analytical Reagents of the American Chemical Society in all tests.
Benzoic Acid - Standard (C6H5COOH) - Pellets made from benzoic acid available from the National Institute of Standards and Technology (NIST) or benzoic acid calibrated against NIST standard material. The calorific value of benzoic acid, for us in the calibration calculations, shall be traceable to a recognized certificate value.
Oxygen - Manufactured from liquid air, guaranteed to be greater than 99.5% pure, and free from combustible matter. Oxygen made by the electrolytic process contains small amounts of hydrogen rendering it unfit unless purified by passage over copper oxide at 500°C.
Titiration Indicator - Methyl orange, methyl red, or methyl purple for indicating the end point when titrating the acid formed during combustion. The same indicator shall be used for both calibration and calorific value determinations.
Standard Solution - Sodium Carbonate (Na2CO3) or other suitable standard solution. A convenient standard solution may be prepared as follows. Dissolve 3.706g of sodium carbonate, dried for 24h at 105°C, in water and dilute to 1L. The resulting concentration (0.0699N) assumes the energy of formation of HNO3 under bomb conditions is -59.7 kJ/mol (-14.3 kcal/mol). One milliliter of this solution is 1.0ml of a 0.10000N base solution is equivalent to 6.0J (1.4 cal) in the acid titration. In general, one milliliter of an arbitrary standard titrant solution is equivalent to its concentration (equivalent/liter or normality [N]) times 59.7J (14.3cal).
HAZARDS
The following precautions are recommended for safe calorimeter operation. Additional precautions are noted in Practice E144. Also consult the calorimeter equipment manufacturer's installation and operating instructions before using the calorimeter.
The mass of sample and any combustion aid as well as the pressure of the oxygen admitted to the bomb shall not exceed the bomb manufacturer's specifications.
Inspect the bomb parts carefully after each use. Replace cracked or significantly worn parts. Replace O-Rings and valve seats in accordance with manufacturer's instruction. For more details, consult the manufacturer.
Equip the oxygen supply cylinder with an approved type of safety device, such as a relief valve, in addition to the needle valve and pressure gauge used in regulating the oxygen feed of the bomb. Valves, gauges, and gaskets shall meet industry safety codes. Suitable reducing valves and adapters for 3- to 4-MPa (300- to 500-psig) discharge pressure can be obtained from commercial sources of compressed gas equipment. Check the pressure gauge annually for accuracy or after any accidental over pressures that reach maximum gauge pressure.
During ignition of a sample, the operator shall not extend any portion of the body over the calorimeter.
Do not fire the bomb if the bomb has been dropped or turned over after loading.
Do not fire the bomb if there is evidence of gas leakage when the bomb is submerged in the calorimeter vessel.
For manually operated calorimeters, the ignition switch shall be depressed only long enough to fire the charge.