ASTM D4809-13 – International Standards

ASTM D4809-13

STANDARD TEST METHOD FOR HEAT OF COMBUSTION OF LIQUID HYDROCARBON FUELS BY BOMB CALORIMETER (PRECISION METHOD)


Scope

This test method covers the determination of the heat of combustion of hydrocarbon fuels. It is designed specifically for use with aviation turbine fuels when the permissible difference between duplicate determinations is of the order of 0.2%. It can be used for a wide range of volatile and nonvolatile materials where slightly greater differences in precision can be tolerated.

In order to attain this precision, strict adherence to all details of the procedure is essential since the error contributed by each individual measurement that affects the precision shall be kept below 0.04%, insofar as possible.

Under normal conditions, the test method is directly applicable to such fuels as gasoline, kerosene, Nos.1 and 2 fuel oils, Nos. 1-D and 2-D diesel fuel and Nos. 0-GT, 1-GT- and 2-GT gas turbine fuels.

Through the improvement of the calorimeter controls and temperature measurements, the precision is improved over that of Test Method D240.

The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.

This standard does not purport to address the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

TERMINOLOGY

Gross Heat of Combustion - expressed as mega joules per kilogram. The gross heat of combustion at constant volume of a liquid or solid fuel containing only the elements carbon, hydrogen, oxygen, nitrogen, and sulfur is the quantity of heat liberated when a unit mass of the fuel is burned in oxygen in an enclosure of constant volume, the products of combustion being gaseous carbon dioxide, nitrgoen, sulfur dioxide, and liquid water, with the initial temperature of the fuel and the oxygen and the final temperature of the product at 25°C. Gross heat of combustion is represented by the symbol Qg3.

Net Heat of Combustion - expressed as mega joules per kilogram. The net heat of combustion at constant pressure of a liquid or solid fuel containing only the elements carbon, hydrogen, oxygen, nitrogen, and sulfur is the quantity of heat liberated when a unit mass of the fuel is burned in oxygen at a constant pressure of 0.101 MPa, the products of combustion being carbon dioxide, nitrogen, sulfur dioxide, and water, all in the gaseous state, with the initial temperature of the fuel and the oxygen and the final temperature of the products of combustion at 25°C.

Temperaturesare measured in degrees Celsius (C).

Timeis expressed in minutes and decimal fractions thereof. It can be measured in minutes or seconds, or both.

Massesare measured in grams. No buoyancy corrections are applied except to obtain the mass of benzoic acid.

The energy unit of measurement employed in this test method is the joule with the heat of combustion reported in mega joules per kilogram.

SUMMARY OF TEST METHOD

The heat of combustion is determined by burning a weighed samples in an oxygen-bomb calorimeter under controlled conditions. The temperature increase is measured by a temperature reading instrument which allows the precision of the test method to be met. The heat of combustion is calculated from temperature observations before, during, and after combustion, with proper allowance for thermo-chemical and heat-transfer corrections. Either isoperibol or adiabatic calorimeters may be used.

SIGNIFICANCE AND USE

The heat of combustion is a measure of the energy available from a fuel. A knowledge of this value is essential when considering the thermal efficiency of equipment for producing either power or heat.

The mass heat of combustion, that is, the heat of combustion per unit mass of fuel, is measured by the procedure. Its magnitude is particularly important to weight-limited vehicles such as airplanes, surface effect vehicles, and hydrofoils as the disturbance such craft can travel on a given weight of fuel is a direct function of the fuel's mass heat of combustion and its density.

The volumetric heat of combustion, that is, the heat of combustion per unit volume of fuel, can be calculated by multiplying the mass heat of combustion by the density of the fuel (mass per unit volume). The volumetric heat of combustion, rather than the mass heat of combustion, is important to volume-limited craft such as automobiles and ships, as it is directly related to the distance traveled between refueling.

APPARATUS AND FACILITIES

Test Room, Bomb, Calorimeter, Jacket, Thermometers and Accessories.

Semi-micro Analytical Balance, having a sensitivity of 0.01mg.

Heavy-Duty Analytical Balance, having a sensitivity of 0.05g.

REAGENTS

Purity of Reagents - Reagent grade chemical shall be used in all tests. Unless otherwise indicated, it is intended that all reagents shall conform to the specifications of the Committee on Analytical Reagents of the American Chemical Society, where such specifications are available. Other grades may be used, provided it is first ascertained that the reagent is of sufficiently high purity to permit its use without lessening the accuracy of the determination.

Purity of Water - Unless otherwise indicated, references to water shall be understood to mean reagent water conforming to Specification D1193, Type IV or better.

Benzoic Acid - The acid must be pelleted before use.

Firing Wire - 0.127mm (No.36 gauge) platinum wire, No.34 gauge iron wire or Chromel C resistance wire, cut in 100m lengths.

Methyl Red Indicator.

Oxygen - Commercial oxygen produced from liquid air can be used without purification (Warning - Oxygen vigorously accelerates combustion). Oxygen prepared by electrolysis of water cannot be used without purification as it can contain some hydrogen. Combustible impurities may be removed by passage over copper oxide at 500C.

Pressure-Sensitive Tape - Cellophane tape 38mm wide, free of chlorine and sulfur.

Alkali, Standard Solutions.

Sodium Hydroxide Solution (0.0866 N) - Dissolve 3.5g of sodium hydroxide (NaOH) in water and dilute to 1L. (Warning - Corrosive. Can cause severe burns or blindness. Evolution of heat produces a violent reaction or eruption upon too rapid mixture with water). Standardize with potassium acid phthalate and adjust to 0.0866 N as described in Practice E200, or alternative use.

Sodium Carbonate Solution (0.0725 N) - Dissolve 3.84g of Na2CO3 in water and dilute to 1L.

2,2,4-Trimethylpentane (0.0866 N) - (isooctane), Standard (Warning - Extremely flammable. Harmful if inhaled. Vapors may cause flash fire)

PREPARATION OF APPARATUS

Arrangement of Apparatus - Install the thermometers as recommended by the manufacturer of the calorimeter. Position the liquid-in-glass thermometer so that the bulb is halfway to the bottom of the bucket and locate the thermistor with its sensing element at about the midpoint of the thermometer bulb. Mount these elements so that exactly the same length is immersed each time the calorimeter is used. Install a thermistor in the water jacket with the elements immersed to the same depth as in the bucket. It is helpful, but not necessary to have liquid-in-glass calorimetric thermometers in both the bucket and jacket for quick temperature observations. Thermistors can be taped to these thermometers. If the thermistors are taped to the thermometers, it can be done in such a manner that the sensing elements are at the midpoint of the thermometer bulbs. The thermometer bulbs and temperature-sensing elements shall not touch the bomb, bucket, or water jacket.

Calorimeter Jacket Controller and Auxiliary Equipment - Adjust the jacket controller, valves, heater, etc. as recommended by the calorimeter manufacturer.

Test Room - The room in which the calorimeter is opened must be free from drafts and not subject to sudden temperature changes. The exact temperature is not important as long as it is in the range from 23 to 26°C and is held constant. The temperature must be constant, not only throughout the day, but from one time of the year to another. The direct rays of the sun shall not strike the calorimeter jacket, bridge, and galvanometer. Adequate facilities for lighting, heating, and ventilation should be provided. Thermostatic control of room temperature and controlled humidity are desirable.

Oxygen Bomb - the oxygen bomb shall meet the requirements specified in Practice E144 and have an internal volume of 350 ± 50ml. All parts shall be constructed of materials that are not affected by the combustion process or products sufficient to introduce measurable heat input or alteration of end products. If the bomb is lined with platinum or gold, all openings shall be sealed to prevent combustion products from reaching the base metal. The bomb must be designed so that all liquid combustion products can be completely recovered by washing the inner surfaces. There must be no gas leakage during a test. The bomb must be capable of withstanding a hydrostatic pressure test to a gauge pressure of 20.7 MPa at room temperature without stressing any part beyond its elastic limit. If necessary, modify the bomb such that the feet are 12mm high to allow for better water circulation under the bomb.

Calorimeter Bucket - The calorimeter bucket shall be made of metal (preferably copper or brass) with a tarnish-resistant coating, and with all outer surfaces highly polished. Its size shall be such that the bomb will be completely immersed in water when the calorimeter is assembled. It shall have a device for stirring the water thoroughly and at a uniform rate but with minimum heat input. Continuous stirring for 10 min shall not raise the temperature more than 0.01°C starting with identical temperatures in the calorimeter, room and jacket. The immersed portion of the stirrer shall be coupled to the outside through a material of low-heat conductivity.

Jacket - The calorimeter bomb, bucket, and water shall be completely enclosed within a stirred water jacket and supported so that its sides, top, and bottom are approximately 1cm from the jacket walls. The jacket may be arranged so as to remain at a constant temperature, or with provisions for automatically adjusting the jacket temperature to equal that of the calorimeter bomb, bucket, and water for adiabatic operation. It must be constructed so that any water evaporating from the jacket will not condense on the calorimeter bucket.

Thermometers - Temperature in the jacket and the calorimeter, respectively, shall be measured with the following thermometers: Etched stem, liquid-in-glass, ASTM Bomb Calorimeter Thermometers, having a range from 19 to 35°C, 18.9 to 25.1°C, or 23.9 to 30.1°C as specified, and conforming to the requirements for Thermometer 56°C, 116°C or 117°C, respectively, as prescribed in Specification E1. Each of these thermometers shall have been tested for accuracy at intervals no larger than 1.5°C over the entire graduated scale. Corrections shall be calculated to 0.002°C, respectively, for each test point. Temperature in the calorimeter shall be measured with the following thermometers: Platinum resistance, thermistor, or other temperature measuring devices which when used with its associated instrumentation will measure the temperature rise repeatably with a recorded resolution of 0.0001°C and a repeatability such that the precision requirements that the mean temperature of all determinations not deviate by more than 0.05°C and that the temperature rise for all experiments be within ± 0.3°C.

Automatic Calorimeter Microprocessor Controller - in place of manually recording temperature and calculating the energy equivalent, an automatic controller may be used. It shall be capable of storing calorimeter temperature readings taken at accurate intervals, firing the bomb and ending the experiment. The stored information along with the sample mass, nitric acid correction and sulfur content, when entered into the instrument, shall be used to calculate the desired energy equivalent or heat of combustion.

Firing Circuit- A 6 to 16V alternating current is required for ignition purposes with an ammeter or pilot light in the circuit to indicate when current is flowing.A step-down transformer connected to a 115V. 50/60-Hz lighting circuit or storage battery may be used (Warning - The ignition circuit switch shall be of the momentary contact type, normally open, except when held closed by the operator).

Timing Device - a watch or other timing device capable of measuring time to 1s is required for use with the isoperibol method.

Sample Cup - A low-form platinum cup, 26mm in diameter and 11mm deep with a spun rim. Base metal alloy crucibles are acceptable if after a few preliminary firings and weight does not change significantly between tests.

STANDARDIZATION

Energy Equivalent of the Calorimeter - Benzoic Acid shall be used as the primary standard (Warning - Oxygen vigorously accelerates combustion). Choose a sample mass so that the temperature rise is approximately equivalent to an energy change of 30,000J. Initially determine the energy equivalent by averaging six determinations made using benzoic acid over a period of at least 3 days.

A relative standard deviation (RSD) of 0.1% or less for the six determinations must be achieved. If not, continue to run until six determinations establish a value that has a RSD of 0.1% or better. If this degree of precision cannot be achieved, review the procedure, critical measurement, mechanical operations and everything that may contribute to scatter in the results. After establishing an energy equivalent value, determine the value at frequent intervals using benzoic acid (every 1 or 2 days of testing) with the average of the last six determinations being used for the energy equivalent as long as the last six determinations have a RSD of 0.1% or less.

WARNING STATEMENTS

Oxygen

  • Do not exceed the sample size limits
  • Follow manufacturer's recommendations for filling the bomb
  • Do not use oil or grease on regulators, gauges, or control equipment
  • Use only with equipment conditioned for oxygen service by carefully cleaning to remove oil grease, and other combustibles
  • Keep combustibles away from oxygen and eliminate ignition sources
  • Keep surfaces clean to prevent ignition or explosion, or both, on contact with oxygen
  • Always use a pressure regulator. Release regulator tension before opening cylinder valve
  • All equipment and containers used must be suitable and recommended for oxygen service
  • Never attempt to transfer oxygen from cylinder in which it is received to any other cylinder
  • Do not mix gases in cylinders
  • Do not drop cylinder. Make sure cylinder is secure
  • Stand away from outlet when opening cylinder valve
  • Keep cylinder out of sun and away from heat
  • Keep cylinder from corrosive environment
  • Do not use cylinder without label
  • Do not use dented or damaged cylinders
  • For technical use only. Do not use for inhalation purposes
  • Use only in well-ventilated area
  • See compressed gas association booklets G-4 and G4.1 for details of safe practice in the use of oxygen.

Sodium Hydroxide

  • Before using, secure information on procedures and protective measures for safe handling
  • Do not get in eyes, on skin, on clothing
  • Avoid breathing dusts or mists
  • Do not take internally
  • When handling, use chemical safety goggles or face shield, protective gloves, boots and clothing
  • When mixing with water, add slowly to surface of solution to avoid violent splattering. In the preparation of solutions do not use hot water, limit temperature rise, with agitation, to 10C/min or limit solution temperature to a maximum of 90C. No single addition should cause a concentration increase greater than 5%.

2,2,4-Trimethylpentane

  • Keep away from heat, sparks, and open flame
  • Keep container closed
  • Use with adequate ventilation
  • Avoid buildup of vapors and eliminate all sources of ignition, especially non-explosion-proof electrical apparatus and heaters.
  • Avoid prolonged breathing of vapor or spray mist
  • Avoid prolonged or repeated skin contact